This document provides instructions for updating the Rigol DP800 series DC powers.

[Update Requirements] 
    - Firmware version: 00.01.16.00.01 and later

[Notices]

    - Before updating, please review the Release Notes for detailed information about the firmware update.

    - Ensure your USB flash drive is functioning properly. This series requires a FAT32-formatted USB flash drive, preferably 16GB or smaller.

    - Do not power off the instrument or remove the USB drive during the update process, as this may cause the device to malfunction.
	
    - User data and calibration values will be retained.
		
[Update Procedure]

1. Notice that the 5 dark gray buttons underneath the unit screen are defined as M1, M2....to M5 from left to right. Press the [Utility] button on the front panel to open the Utility menu. Select Sys Info and press M1 -> M3 -> M2 to view system details. Confirm the instrument meets the update requirements. If it does, proceed. If not, do not continue — an incompatible update may cause critical issues, including start failure and device damages.
  
2. Extract the firmware package and copy the GEL file to the root directory of the USB drive. Do not rename the file.
  
3. Power off the instrument, then insert the USB drive into the USB port at the back of the instrument.
  
4. Power on the instrument. When the RIGOL logo appears with three dots underneath, immediately press the [Help] button. The update process bar will appear, indicating that the update has started.

5. Once the update finishes, the instrument will proceed to the waveform configuration interface. Remove the USB flash drive.

6. Press the [Help] button twice to open the Help menu. Then press these buttons in order: M4 -> M2 -> M1 to update analog board 1; M4 -> M2 -> M2 to update analog board 2.

7. Restart the instrument.

8. Check whether the firmware has been updated: navigate to [Utility] -> Sys Info, and then press M1 -> M3 -> M2 to check.
  
  
[troubleshooting]

    - If the instrument fails to read the update file:
      try using a different USB flash drive.

    - If the issue persists:
      re-download the firmware package and repeat the update steps.

    - If the problem continues after the above steps:
      please contact your regional support team for assistance.

· Africa, Middle East & Southern Asia: info.in@rigol.com
· Canada & the US: help@rigol.com
· China: info.cn@rigol.com
· Europe: info-europe@rigol.com
· Japan: info.jp@rigol.com
· Korea: info.kr@rigol.com
· South America: info.br@rigol.com
· All Other Regions: info.int@rigol.com